Implant Extraction System

The first system on the market patented by BTI for safe extraction.

The only system on the market that guarantees atraumatic extraction of internal and external connection implants, in any situation.

Extractor kit

Five exclusive extractors indicated for the extraction of osseointegrated implants when they are no longer functional or are unnecessary in a new prosthetic situation.


Implante Externo 2

An easy and precise procedure

Extraccion_Implantes1 Extraccion_Implantes2 Extraccion_Implantes3 Extraccion_Implantes4

Choose the appropriate implant extractor. Choose the right extractor for the shape of the internal and external implant connections.

Manually insert the extractor in the implant thread, using the handle-adapter, ensuring it fits in an anti-clockwise direction.

Connect the counter-torque wrench and turn it anti-clockwise.

Apply torque gently in an anti-clockwise direction until the implant is fully extracted.

Counter-Torque Wrench

Wrench with large torque that breaks the join between the bone and impact, making the extraction process easier. When torque over 200 Ncm is applied, the wrench forms an angle and prevents progress. Back to the original position using a cylindrical component in the kit.

MDPT

Handle to carry the extractor once the extender has been fitted.

 

Extnders

Set of counter-torque wrench extenders(long, short, extralong) for different clinical situations.

 

Extractors

Five parts with varying widths and lengths to adapt to the range of implants on the market. If necessary due to the depth of the connection, they can be cut to size using a diamond disc cutter. They are disposable to ensure maximum performance.

Trephine Drills

An atraumatic trephine drill kit for an ultrafine cut to release the area with maximum osseointegration, preserving the surrounding bone, only drilling the first 3 millimetres.

Trephine drilling to just 2-3mm reduces maximum torque to less than half, to break the implant-bone connection with guaranteed extraction success.

They allow an ultrafine cut along the first 5 millimetres because they are narrower at that level, preserving the surrounding bone. This is particularly important in the case of resorbed maxilla.

They are delivered in four different diameters, depending on the width of the implants