Assessing the Impact of IOS Scanning Accuracy on Additively Manufactured Occlusal Splints

Abstract

Introduction: Digital workflow and intraoral scanners (IOSs) are used to clinically obtain data for a wide range of applications in restorative dentistry. The study aimed to compare two different IOSs with inexperienced users in the digital workflow of oral split manufacturing.

Material and Methods: Anonymous stone models of upper and lower dentate patients were used. Both models were scanned with a desktop 3D scanner 3Shape D2000 to obtain the reference models (STLR). Ten inexperienced operators scanned each model three times with each IOS system (3Shape TRIOS 3 and Carestream CS 3800). Finally, 20 intraoral scanners were randomly chosen from the obtained dataset (10 per IOS system) to design and manufacture 20 nightguards. All the nightguards were scanned. Trueness and precision were calculated and compared between the two IOS systems.

Results: All the mean errors both for trueness and precision were below 40 µm, more than acceptable for the design and manufacturing of intraoral devices such as nightguards. All the mean errors (except one) for trueness between the inner part of the nightguards and the upper control model were below 100 µm, less than a printed layer height. For inexperienced operators, both IOSs are suitable for a digital workflow of manufacturing occlusal splints.

Keywords: accuracy; digital dentistry; intraoral scanner; nightguard; precision; trueness.

Figures

— actualidad · noticias ›

BTI Biotechnology Institute supera una década de liderazgo científico entre las bioempresas españolas

miércoles, 10 de junio de 2026

BTI Biotechnology Institute vuelve a situarse como la empresa biotecnológica española con mayor producción científica, con un total de 29 publicaciones recogidas en el Informe AseBio 2025.

Dent J (Basel)

2,024

Autores:
Lazcano A,Alkhraisat MH.

Abstract

Introduction: Digital workflow and intraoral scanners (IOSs) are used to clinically obtain data for a wide range of applications in restorative dentistry. The study aimed to compare two different IOSs with inexperienced users in the digital workflow of oral split manufacturing.

Material and Methods: Anonymous stone models of upper and lower dentate patients were used. Both models were scanned with a desktop 3D scanner 3Shape D2000 to obtain the reference models (STLR). Ten inexperienced operators scanned each model three times with each IOS system (3Shape TRIOS 3 and Carestream CS 3800). Finally, 20 intraoral scanners were randomly chosen from the obtained dataset (10 per IOS system) to design and manufacture 20 nightguards. All the nightguards were scanned. Trueness and precision were calculated and compared between the two IOS systems.

Results: All the mean errors both for trueness and precision were below 40 µm, more than acceptable for the design and manufacturing of intraoral devices such as nightguards. All the mean errors (except one) for trueness between the inner part of the nightguards and the upper control model were below 100 µm, less than a printed layer height. For inexperienced operators, both IOSs are suitable for a digital workflow of manufacturing occlusal splints.

Keywords: accuracy; digital dentistry; intraoral scanner; nightguard; precision; trueness.

Figures

Artículos relacionados

Continúa tu investigación a través de nuestra selección de publicaciones vinculadas. Descubre estudios complementarios que abordan áreas temáticas, enfoques clínicos o desarrollos afines, diseñados para enriquecer tu comprensión sobre este campo científico.

LESIONES NERVIOSAS

The Effectiveness of Plasma Rich in Growth Factors in the Treatment of Nerve Compression Syndromes of the Upper Extremity

Galan V, Inigo-Dendariarena I, Galan J, Prado R, Padilla S, Anitua E. J Clin Med. 2022 Aug 16;11(16):4789.

LESIONES NERVIOSAS

The Effectiveness of Plasma Rich in Growth Factors in the Treatment of Nerve Compression Syndromes of the Upper Extremity

Galan V, Inigo-Dendariarena I, Galan J, Prado R, Padilla S, Anitua E. J Clin Med. 2022 Aug 16;11(16):4789.

LESIONES NERVIOSAS

The Effectiveness of Plasma Rich in Growth Factors in the Treatment of Nerve Compression Syndromes of the Upper Extremity

Galan V, Inigo-Dendariarena I, Galan J, Prado R, Padilla S, Anitua E. J Clin Med. 2022 Aug 16;11(16):4789.

Contacta

Primo piano
Nombre *

Apellidos *

Correo electrónico *

¿Cómo quieres que contactemos contigo? *

Teléfono *

Código postal *

País *

Escribe aquí tu consulta *